Clinical Usefulness of the Active Fixation Quadripolar Left Ventricular Lead Compared With the Passive Fixation Quadripolar Lead in Cardiac Resynchronization Therapy.
Cardiac resynchronization therapy
Left ventricular lead-related complications
Quadripolar active fixation lead
Journal
Circulation journal : official journal of the Japanese Circulation Society
ISSN: 1347-4820
Titre abrégé: Circ J
Pays: Japan
ID NLM: 101137683
Informations de publication
Date de publication:
04 Jul 2024
04 Jul 2024
Historique:
medline:
4
7
2024
pubmed:
4
7
2024
entrez:
3
7
2024
Statut:
aheadofprint
Résumé
This study compared the stability of the Medtronic Attain Stability Quad (ASQ), a novel quadripolar active fixation left ventricular (LV) lead with a side helix, to that of conventional quadripolar leads with passive fixation (non-ASQ) and evaluated their LV lead performance.Methods and Results: In all, 183 consecutive patients (69 ASQ, 114 non-ASQ) who underwent cardiac resynchronization therapy (CRT) between January 2018 and June 2021 were enrolled. Complications, including elevated pacing capture threshold (PCT) levels, phrenic nerve stimulation (PNS), and LV lead dislodgement, were analyzed during the postimplantation period until the first outpatient visit after discharge. The frequency of LV lead-related complications was significantly lower in the ASQ than non-ASQ group (14% vs. 30%, respectively; P=0.019). Specifically, LV lead dislodgement occurred only in the non-ASQ group, and elevated PCT levels were significantly lower in the ASQ group (7% vs. 23%; P=0.007). Kaplan-Meier analysis confirmed a significantly lower incidence of LV lead-related complications in the ASQ group (log-rank P=0.005). Cox multivariable regression analysis showed a significant reduction in lead-related complications associated with ASQ (hazard ratio 0.44; 95% confidence interval 0.23-0.83; P=0.011). The ASQ group exhibited fewer LV lead-related complications requiring reintervention and setting changes than the non-ASQ group. Thus, the ASQ may be a favorable choice for CRT device implantation.
Sections du résumé
BACKGROUND
BACKGROUND
This study compared the stability of the Medtronic Attain Stability Quad (ASQ), a novel quadripolar active fixation left ventricular (LV) lead with a side helix, to that of conventional quadripolar leads with passive fixation (non-ASQ) and evaluated their LV lead performance.Methods and Results: In all, 183 consecutive patients (69 ASQ, 114 non-ASQ) who underwent cardiac resynchronization therapy (CRT) between January 2018 and June 2021 were enrolled. Complications, including elevated pacing capture threshold (PCT) levels, phrenic nerve stimulation (PNS), and LV lead dislodgement, were analyzed during the postimplantation period until the first outpatient visit after discharge. The frequency of LV lead-related complications was significantly lower in the ASQ than non-ASQ group (14% vs. 30%, respectively; P=0.019). Specifically, LV lead dislodgement occurred only in the non-ASQ group, and elevated PCT levels were significantly lower in the ASQ group (7% vs. 23%; P=0.007). Kaplan-Meier analysis confirmed a significantly lower incidence of LV lead-related complications in the ASQ group (log-rank P=0.005). Cox multivariable regression analysis showed a significant reduction in lead-related complications associated with ASQ (hazard ratio 0.44; 95% confidence interval 0.23-0.83; P=0.011).
CONCLUSIONS
CONCLUSIONS
The ASQ group exhibited fewer LV lead-related complications requiring reintervention and setting changes than the non-ASQ group. Thus, the ASQ may be a favorable choice for CRT device implantation.
Identifiants
pubmed: 38960680
doi: 10.1253/circj.CJ-24-0084
doi:
Types de publication
Journal Article
Langues
eng
Sous-ensembles de citation
IM