End-to-end testing for stereotactic radiotherapy including the development of a Multi-Modality phantom.
Dynamic Tumor Tracking
End-to-End testing
MRI distortion
Quality Assurance
SBRT
SRS
Journal
Zeitschrift fur medizinische Physik
ISSN: 1876-4436
Titre abrégé: Z Med Phys
Pays: Germany
ID NLM: 100886455
Informations de publication
Date de publication:
18 Dec 2022
18 Dec 2022
Historique:
received:
24
06
2022
revised:
16
11
2022
accepted:
16
11
2022
entrez:
20
12
2022
pubmed:
21
12
2022
medline:
21
12
2022
Statut:
aheadofprint
Résumé
A new insert for a commercially available end-to-end test phantom was designed and in-house manufactured by 3D printing. Subsequently, the insert was tested for different stereotactic radiation therapy workflows (SRS, SBRT, FSRT, and Multimet) also in comparison to the original insert. Workflows contained imaging (MR, CT), treatment planning, positioning, and irradiation. Positioning accuracy was evaluated for non-coplanar x-ray, kV- and MV-CBCT systems, as well as surface guided radiation therapy. Dosimetric accuracy of the irradiation was measured with an ionization chamber at four different linear accelerators including dynamic tumor tracking for SBRT. CT parameters of the insert were within the specification. For MR images, the new insert allowed quantitative analysis of the MR distortion. Positioning accuracy of the phantom with the new insert using the imaging systems of the different linacs was < 1 mm/degree also for MV-CBCT and a non-coplanar imaging system which caused > 3 mm deviation with the original insert. Deviation of point dose values was <3% for SRS, FSRT, and SBRT for both inserts. For the Multimet plans deviations exceeded 10% because the ionization chamber was not positioned in each metastasis, but in the center of phantom and treatment plan. The in-house manufactured insert performed well in all steps of four stereotactic treatment end-to-end tests. Advantages over the commercially available alternative were seen for quantitative analysis of deformation correction in MR images, applicability for non-coplanar x-ray imaging, and dynamic tumor tracking.
Identifiants
pubmed: 36539322
pii: S0939-3889(22)00123-4
doi: 10.1016/j.zemedi.2022.11.006
pii:
doi:
Types de publication
Journal Article
Langues
eng
Sous-ensembles de citation
IM
Informations de copyright
Copyright © 2022 The Authors. Published by Elsevier GmbH.. All rights reserved.