Intracranial stereotactic radiotherapy in off-isocenter target with SyncTraX FX4.
Localization accuracy
Off-isocenter target
Stereotactic radiotherapy
SyncTraX
Journal
Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB)
ISSN: 1724-191X
Titre abrégé: Phys Med
Pays: Italy
ID NLM: 9302888
Informations de publication
Date de publication:
Aug 2022
Aug 2022
Historique:
received:
20
12
2021
revised:
13
06
2022
accepted:
29
06
2022
pubmed:
8
7
2022
medline:
10
8
2022
entrez:
7
7
2022
Statut:
ppublish
Résumé
We investigated the localization accuracy of the off-isocenter targets using SyncTraX FX4, a new image registration device. In a phantom study, we used a MultiMet-WL Cube with metal targets at different distances from the isocenter. Image registrations were performed with SyncTraX and cone-beam computed tomography (CBCT). Nineteen fields with different gantry, collimator, and couch angles were delivered to each target. Localization errors of the off-isocenter targets were then evaluated. In a clinical study, localization accuracy was evaluated for 32 patients. First, image registration was performed using SyncTraX, and the accuracy of patient positioning was evaluated using CBCT. Next, positioning corrections were performed for intracranial setup errors exceeding the threshold (0.5 mm/0.5°) in each field. Finally, total setup uncertainty was evaluated using CBCT. Differences in dosimetric errors from planned doses between no patient positioning corrections during treatment and positioning corrections with SyncTraX were also evaluated. In the phantom study, the positioning accuracy on targets up to 7 cm from the isocenter was within 1 mm. In the clinical practice, the localization accuracies of SyncTraX were 0.35 ± 0.39 mm, 0.30 ± 0.24 mm, and 0.03 ± 0.27 mm in the lateral, vertical, and longitudinal directions, respectively. Post-treatment setup errors were reduced by correcting intrafractional setup errors with SyncTraX during treatment. Positioning corrections with SyncTraX reduced the maximum dosimetric error from 1.6% to 1.0%. SyncTraX provides satisfactory localization accuracy for the off-isocenter targets within 7 cm. SyncTraX reduce dosimetric errors caused by intrafractional setup errors during treatment.
Identifiants
pubmed: 35797917
pii: S1120-1797(22)02012-9
doi: 10.1016/j.ejmp.2022.06.020
pii:
doi:
Types de publication
Journal Article
Langues
eng
Sous-ensembles de citation
IM
Pagination
105-111Informations de copyright
Copyright © 2022 Associazione Italiana di Fisica Medica e Sanitaria. Published by Elsevier Ltd. All rights reserved.