Retrospective four-dimensional magnetic resonance imaging of liver: Method development.
SBRT
liver
radiotherapy
retrospective 4D-MRI
Journal
Journal of applied clinical medical physics
ISSN: 1526-9914
Titre abrégé: J Appl Clin Med Phys
Pays: United States
ID NLM: 101089176
Informations de publication
Date de publication:
Dec 2020
Dec 2020
Historique:
received:
26
06
2020
revised:
29
10
2020
accepted:
06
11
2020
pubmed:
4
12
2020
medline:
22
6
2021
entrez:
3
12
2020
Statut:
ppublish
Résumé
Purpose of our research was to develop a four-dimensional (4D) magnetic resonance imaging (MRI) method of liver. Requirements of the method were to create a clinical procedure with acceptable imaging time and sufficient temporal and spatial accuracy. The method should produce useful planning image sets for stereotactic body radiation therapy delivery both during breath-hold and in free breathing. The purpose of the method was to improve the localization of liver metastasis. The method was validated with phantom tests. Imaging parameters were optimized to create a 4D dataset compressed to one respiratory cycle of the whole liver with clinically reasonable level of image contrast and artifacts. Five healthy volunteers were imaged with T2-weighted SSFSE research sequence. The respiratory surrogate signal was observed by the linear navigator interleaved with the anatomical liver images. The navigator was set on head-feet - direction on the superior surface of the liver to detect the edge of diaphragm. The navigator signal and 2D liver image data were retrospectively processed with a self-developed MATLAB algorithm. A deformable phantom for 4D imaging tests was constructed by combining deformable tissue-equivalent material and a commercial programmable motor unit of the 4D phantom with a clinically relevant range of deformation patterns. 4D Computed Tomography images were used as reference to validate the MRI protocol. The best compromise of reasonable accuracy and imaging time was found with 2D T2-weighted SSFSE imaging sequence using parameters: TR = 500-550 ms, images/slices = 20, slice thickness = 3 mm. Then, image processing with number of respiratory phases = 8 constructed accurate 4D images of liver. We have developed the 4D-MRI method visualizing liver motions three-dimensionally in one representative respiratory cycle. From phantom tests it was found that the spatial agreement to 4D-CT is within 2 mm that is considered sufficient for clinical applications.
Identifiants
pubmed: 33270997
doi: 10.1002/acm2.13108
pmc: PMC7769409
doi:
Types de publication
Journal Article
Langues
eng
Sous-ensembles de citation
IM
Pagination
304-313Informations de copyright
© 2020 The Authors. Journal of Applied Clinical Medical Physics published by Wiley Periodicals LLC on behalf of American Association of Physicists in Medicine.
Références
Int J Radiat Oncol Biol Phys. 2017 Sep 1;99(1):227-237
pubmed: 28816151
Radiother Oncol. 2014 May;111(2):252-7
pubmed: 24746577
Radiother Oncol. 2016 Nov;121(2):276-280
pubmed: 27825795
Med Phys. 2014 Oct;41(10):101902
pubmed: 25281954
Radiother Oncol. 2018 Dec;129(3):486-493
pubmed: 29871813
Phys Med Biol. 2015 Nov 21;60(22):R323-61
pubmed: 26509844
J Magn Reson Imaging. 2014 May;39(5):1301-7
pubmed: 24719241
Med Phys. 2004 Mar;31(3):427-32
pubmed: 15070239
Phys Med Biol. 2016 May 7;61(9):3472-87
pubmed: 27049817
Phys Med Biol. 2007 Mar 21;52(6):1547-64
pubmed: 17327648
Med Phys. 2015 Aug;42(8):4460-71
pubmed: 26233176
Int J Radiat Oncol Biol Phys. 2008 Jul 1;71(3):907-15
pubmed: 18514783
Int J Radiat Oncol Biol Phys. 2015 Mar 1;91(3):571-8
pubmed: 25596109
Med Phys. 2007 Sep;34(9):3620-9
pubmed: 17926966
Magn Reson Imaging. 2012 Nov;30(9):1323-41
pubmed: 22770690
Phys Med Biol. 2018 Oct 23;63(21):21TR01
pubmed: 30272573
Radiother Oncol. 2018 Jun;127(3):474-480
pubmed: 29804801
Med Phys. 2003 Oct;30(10):2762-92
pubmed: 14596315
Magn Reson Med. 1996 Jul;36(1):117-23
pubmed: 8795030
Radiat Oncol. 2018 Feb 13;13(1):26
pubmed: 29439707
World J Radiol. 2010 Aug 28;2(8):309-22
pubmed: 21160685