Simulation and experimental validation of a prototype electron beam linear accelerator for preclinical studies.
Electron dosimetry
FLASH
Geant4
Monte-Carlo
Journal
Physica medica : PM : an international journal devoted to the applications of physics to medicine and biology : official journal of the Italian Association of Biomedical Physics (AIFB)
ISSN: 1724-191X
Titre abrégé: Phys Med
Pays: Italy
ID NLM: 9302888
Informations de publication
Date de publication:
Apr 2019
Apr 2019
Historique:
received:
28
11
2018
revised:
12
03
2019
accepted:
17
03
2019
entrez:
20
4
2019
pubmed:
20
4
2019
medline:
21
6
2019
Statut:
ppublish
Résumé
Measurements and Monte-Carlo simulations were carried out to model the dose distribution of a prototype electron beam linear accelerator (Kinetron LINAC) to determine the dose to organs in small animal irradiations experiments. Dose distributions were simulated using the GATE8.0/Geant4.10.3 Monte-Carlo platform, and measured in air and solid water phantoms using a commercial scintillating screen detector and new EBT-XD Gafchromic films. The LINAC is able to produce 4.5 MeV electron beams at dose-rates ranging from Gy/min to thousands of Gy/s, and is used to study the radiobiological effects of very-high dose-rates that have been shown to protect normal tissues from radiation toxicity. Numerical simulations and experimental dosimetric characterisation of this electron accelerator were performed with the Monte-Carlo toolkit and various detectors. Absolute dose distributions in solid water were measured and compared with simulations. Realistic electron irradiation conditions were simulated in voxelised mice CT images. 3D dose distributions and dose-volume histograms in lungs of mice were simulated and analyzed. Measured and calculated depth-dose profiles for several beam configurations (energy and dose-rate) were compared. Beam emittance was validated by comparing measured and calculated beam sizes along the central axis in air: the deviation for all conditions was less than 1 mm. A good agreement was obtained between experimental dose distributions and the results obtained with simulations (<2% dose differences for lateral and depth-dose profiles). The method presented here, relying on few free parameters, can be adapted to very-high dose-rate electron irradiation to support the analysis of preclinical research experiments.
Identifiants
pubmed: 31000086
pii: S1120-1797(19)30056-0
doi: 10.1016/j.ejmp.2019.03.016
pii:
doi:
Substances chimiques
Water
059QF0KO0R
Types de publication
Journal Article
Validation Study
Langues
eng
Sous-ensembles de citation
IM
Pagination
50-57Informations de copyright
Copyright © 2019 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.